Best Fitted Sphere 10mm Float Versus Best Fitted Sphere 3mm Centered on the Thinnest Point Measurements; Accuracy for the Diagnosis of Keratoconus
Mahmood Joshaghani1,
Mohammad Soleimani2*, Pirouzeh Farsi3, Khalil Ghasemi
Falavarjani1, Hassan Ghasemi3, Reza Soudi2,
Zahra Joshaghani5, Alireza Foroutan1, Mansour Mozafarri2,
Seyed Ali Tabatabaei2, Zahra Fallah Tafti6
1Eye Research Center, Rassoul Akram Hospital, Iran University of
Medical Sciences, Tehran, Iran
2*Ocular Trauma and Emergency Department, Farabi Eye Hospital,
Tehran University of Medical Sciences, Iran
3Eye Research Center, Rassoul Akram Hospital, Iran University of
Medical Sciences, Tehran, Iran
4Shahed University, Tehran, Iran
5Azad University of Medical Sciences, Tehran, Iran
6Optometrist, Razi Eye Clinic, Tehran, Iran
*Corresponding
author: Mohammad Soleimani,
Anterior Segment Subspecialist. Ocular Trauma and Emergency Department, Farabi
Eye Hospital, Tehran University of Medical Sciences, Iran. Tel: + 0098 912
1096496; Email: Soleimani_md@yahoo.co
Citation:
Joshaghani M, Soleimani M, Farsi M, Falavarjani KG, Ghasemi H, et al. (2017)
Best Fitted Sphere 10mm Float Versus Best Fitted Sphere 3mm Centered on the
Thinnest Point Measurements; Accuracy for the Diagnosis of Keratoconus.
Ophthalmol Res Rep 2017: J121. DOI: 10.29011/ORRT-121. 100021
Purpose: To evaluate the accuracy of 3 mm Best Fitted Sphere (BFS)
centered on the thinnest point measurement to detect keratoconus and to compare
this to BFSs with 10 mm float measurement.
Setting: Eye Research Center, Rassoul Akram Hospital, Iran University of
Medical Sciences, Iran
Materials and Methods: In this cross sectional study, 42 eyes from 29 patients with
confirmed keratoconus were compared with 49 normal eyes of 27 candidates of
refractive surgery.
Anterior and posterior corneal surfaces were analyzed using Orb
scanIIz slit-scanning topography in both groups. BFSs with 10 mm float
measurement and 3mm centered on the thinnest point measurement were compared.
Results: For detecting keratoconus, higher overall predictive accuracies
(area under the curve =100%) for anterior and posterior BFS 3mm with center of
the thinnest point compared to the anterior (98%) and posterior (99.5%) BFS 10
mm float measurements were achieved. There are no cutoff point using BFS 10 mm
measurements for anterior float (range 42-43 diopters) and posterior float
(range 52-54) for discriminating keratoconus from normal eyes. However, with a
cutoff point of 45.4 diopters and 62.75 diopters for anterior and posterior BFS
3mm measurements, respectively, neither normal eyes, nor keratoconus eyes were
misclassified.
Conclusion: Compared to the BFS 10 mm float measurement, BFS 3 mm centered
on the thinnest point effectively discriminates keratoconus from normal
corneas.
Keywords: Best Fitted Sphere; Keratoconus; Orb Scan
1. Introduction
Keratoconus (KCN) is a bilateral non inflammatory corneal
ectasia, it is characterized by a progressive increase in corneal curvature,
with apical thinning and irregular corneal astigmatism [1,2]. Keratorefractive
surgeries are widely used to correct the refractive error. However, any
preexisting tendency to ectasia like keratoconus will be manifested and
aggravated by keratorefractive surgeries [2,3]. Thus the importance of
recognizing such preexisting tendency towards keratectasia cannot be
overemphasized. It is not only important to diagnose these cases of keratoconus
that are clinically evident, but at the same time to identify those subclinical
cases whose cornea is clinically not keratoconic but are detectable by imaging
studies.
The Orb scan slit scanning topography system can provide useful
and accurate information in detecting subtle topographic changes present in
early keratoconus and documenting their progression by serial topographic
analysis [4-6]. For refractive surgery screening, analyzing the elevation data
using a Best Fitted Sphere (BFS) is the most widely used and convenient (i.e.
easiest to read and understand). The instrument is routinely set on BFS 10 mm
float. Using this setting, various parameters were suggested for detecting KCN
[5-8].
Changes in corneal topography in most cases of keratoconus (i.e.
cone) start in infer temporal Para center of cornea and it corresponds to the
thinnest point of the cornea [1]. The importance of this study is that BFS
measurements centered on this location and with lower diameter may enhance the
protrusion of the cone from BFS and discriminate keratoconus from normal
corneas more effectively. The aim of this study was to compare the BFS 10mm
float and BFS 3mm centered on the thinnest point measurements in keratoconus
and normal eyes. To our knowledge, this is the first study that compares orb
scan indices in different BFS measurements in keratoconic and normal eyes.
2. Materials and
Methods
In this cross sectional comparative study, 42 eyes from 29
patients with confirmed keratoconus (KCN group) and 49 control eyes of 27
normal subjects were enrolled. KCN group had clinical signs of keratoconus
including cone and vogtstriae in slit lamp bio microscopy. The study was done
in Rassoul Akram Hospital, Iran University of Medical Sciences in Tehran, Iran
during 2009.
Also the KCN eyes had typical topographic criteria defined for
diagnosis of keratoconus.2Orbscan
inclusion criteria for control group were absence of focal or inferior
steepening of the cornea, Mean sim k < 47.2 D , inferior-superior difference
< 1.4 , thinnest point > 470 mm , Irregularity 3mm < 1.5 , Anterior
BFS measurement < 45 D, Posterior BFS measurement < 52 ,
Anterior/posterior BFS measurements < 1.21 , Ant elevation measurement <
25 , Post elevation measurement < 35 , absence of non orthogonal, asymmetric
and broken bow tie astigmatisms. Exclusion criteria for both groups were
corneal scarring, any signs or history of other corneal disease, history of
previous ocular surgery or trauma, and history of contact lens usage in last 3
weeks. All patients underwent a cycloplegic refraction, slit lamp examinations
and corneal imaging using orb scan (Orb scan II, Bausch & Lomb Inc.,
Rochester, NY, USA).
Recorded variables were anterior and posterior dioptric powers,
anterior and posterior BFS 10 mm float measurements, anterior and posterior BFS
3 mm centered on thinnest point measurements, anterior and posterior corneal
elevations, tachymetry of thinnest point, magnitude of the decent ration of the
thinnest corneal point from the corneal geometric center (X,Y), mean
keratometry (K), sim k (min and max), irregularity 3mm and 5mm, difference of
the thinnest point tachymetry and thickest point in center of 7mm,
inferior/superior difference of and mean astigmatism of the eyes. All of the
orb scan measurements were performed by one of the two experienced technicians
using an acquisition protocol recommended by the manufacturer. First, the
default settings for BFS were used in all cases: floating alignment and full
cornea fit zone (10 mm). The center of all maps was the apex determined by
placid data. Then, the BFS with the diameter of 3mm was centered on the
thinnest point using the “surface rotation” menu by clicking on the “tools”
button on the main toolbar.
Data were analyzed using the SPSS software version 16 (IBM
Corporation). The P values comparing the differences in variables between the
groups of keratoconus and normal eyes were considered significant when values
were less than 0.05. Three trials were used to calculate SD. The
Kolmogorov-Smirnov test was employed to test the normal distribution of all
continuous variables. Two samples independent t test was performed
for normally distributed variables, and nonparametric tests were used if
variables were not normally distributed. Y.S.
Rabinowitz and K. Rasheed, KISA% index: a quantitative videokeratography
algorithm embodying minimal topographic criteria for diagnosing keratoconus, J
Cataract Refract Surg 25 (1999),
pp. 1327–1335 errata 2000; 26:480. Article | PDF (474 K) | View Record in Scopus | Cited By in Scopus
(66) Receiver Operating Characteristic (ROC) curves were used to determine the
test's overall predictive accuracy (area under the curve) and to identify
optimal corneal elevation cutoff points to maximize sensitivity and specificity
in discriminating keratoconus from normal corneas.
3. Results
Twenty-seven normal subjects including 12 males (44.45%) and 15
females (55.55%), and 29 KCN patients including 15 males (51.73%) and 14
females (48.27%) were examined.
Table 1 shows orb scan parameters in normal and keratoconus
eyes. There were significant differences between the normal and keratoconus
eyes in all recorded indices including BFS 10mm float and BFS 3mm centered on
thinnest point measurements.
There was significant difference between the BFS 10 mm float and
BFS 3mm centered on thinnest point measurements in keratoconus and normal eyes
(table 2,3). ROC curves analyses (table 4) shows.
Higher overall predictive accuracy of anterior and posterior BFS
3mm measurements centered on thinnest point (area under the curve =100%)
compared to BFS 10 mm float measurements (area under the curve in anterior
measurement = 98% and in posterior measurement = 99.5%). Using the anterior and
posterior BFS 3mm measurements with a cutoff point of 45.4 and 62.75 diopters,
respectively, neither normal eyes nor keratoconus eyes were misclassified.
However, there was no cutoff point using BFS 10 mm float for discriminating
keratoconus from normal eyes.
4. Discussion
Findings of the present study suggest that anterior and
posterior BFS 3mm centered on the thinnest point effectively discriminates
keratoconus from normal corneas and may detect subclinical keratoconus with
more accuracy than routine BFS 10mm float measurements.
We defined a new reference surface showed that a BFS with 3 mm
diameter and centered on the thinnest portion of the cornea will exaggerate the
cone and results in a significantly steeper BFS, making discrimination of the
abnormal cornea easier. The new BFS shows that the conical portion of the
cornea is now more pronounced. Previous studies reported quantitative indices
and diagnostic criteria for detecting keratoconus using different imaging tools
[4-17]. Our findings are in line with previous studies that have shown various
orb scan parameters are significantly different between the normal and
keratoconus eyes. Lim et al [5]. reported that the mean values of maximum
posterior elevation, thinnest corneal tachymetry, 3-mm irregularity, and 5-mm
irregularity were significantly different between the controls, keratoconic,
and the suspected keratoconus group. In both groups, the indices showed
deviation from the normal ranges, although there was some degree of overlap. It
was reported that anterior and posterior corneal curvatures were affected in
keratoconus and keratoconus suspects [1-4] Rao and associates [6] reported a
high sensitivity and specificity of 91% and 99% for thickness index [7]. showed
that the mean values of irregularity index in 3 mm and 5 mm zones and the
thinnest optical tachymetry were significantly different between the normal
eyes and keratoconic eyes [9]. report that a posterior elevation of 40 micron
or more has sensitivity of 57.7% and specificity of 89.9%, and suggest an
anterior elevation ratio of 0.5122 mm or less had 99% sensitivity and 95.2%
specificity while a ratio 16.5 mm or less had 80.1% sensitivity and 80.8%
specificity in discriminating normal eyes from keratoconus and keratoconus
suspects. Some studies reported special variables for differentiating KCN eyes
without reporting the sensitivity and specifity [6]. Showed that a maximum
central posterior elevation of 40 μm or more is a risk factor for formefruste
keratoconus [5]. Showed that this value should be 46 μm. Prakash and Agarwal
[17] suggested that anterior corneal elevation more than 25 μm and posterior
corneal elevation more than 50 μm are indicative of early keratoconus.
Controversies in the cut off points and criteria in these reports and those
reported by others with orb scan and other instruments [10-17] shows that there
is a large space for further research in this field. We showed that BFS 3mm
centered on thinnest point measurements has 100 % sensitivity and specifity for
discriminating the KCN eyes. Although the anterior and posterior elevations
were significantly different between normal and keratoconus eyes, we couldn’t
find a specific cut off point for these variables. Since the BFS diameter was
decreased, the BFS radius was increased and so the elevations were less
evident. This is in contrast to the Belin et al [13] a concept that believes
the exclusion of a circle from the central corneal 4 mm around the thinnest
point enhances the ectasia. We scarified the elevations instead of BFS
measurements and found it useful.
Our study has some limitations. The sample size of this study
was small. Also, our values for elevations are different from other studies;
this may result from specific inclusion criteria. We included patients with
apparent clinical KCN and the result may be different if patients with
subclinical KCN could be considered. However the results may help to determine
corneal ectasia, the dreadful complication of refractive surgery. We didn’t
evaluate the dioptric corneal power. The average dioptric measurement over the
thinnest 3mm of the cornea may have the same accuracy as the 3mm BFS.
5. Conclusion
In
conclusion, the finding of our study showed that posterior BFS 3mm centered on
thinnest point was significantly different in keratoconus than normal eyes.
There was significantly difference between BFS 10mm float and BFS 3mm centered
on thinnest point of cornea in keratoconus and normal eyes. ROC curves analyses
showed higher overall predictive accuracy of posterior BFS 3mm centered on
thinnest point for keratoconus. We examined BFS measurements in eyes with
apparent clinical KCN. Future larger studies comparing the measurements in form
frost KCN is needed to confirm our findings.
|
Normal eyes |
Keratoconus eyes |
P |
||
|
Mean |
SD |
Mean |
SD |
Value |
BFS Ant 3 T (mm) |
7.95 |
0.17 |
6.16 |
0.63 |
< .0001 |
BFS Ant 3 T (D) |
42.49 |
0.91 |
55.32 |
5.63 |
< .0001 |
BFS Post 3 T (mm) |
6.2 |
0.21 |
4.24 |
0.52 |
< .0001 |
BFS Post 3 T (D) |
54.56 |
1.9 |
80.34 |
10.1 |
< .0001 |
Elevation Ant 3 T (mm) |
0.001 |
0.002 |
0.005 |
0.012 |
< .0001 |
Elevation Post 3 T (mm) |
0.003 |
0.006 |
0.01 |
0.004 |
< .0001 |
BFS Ant 10 F (mm) |
8.12 |
0.15 |
7.36 |
0.41 |
< .0001 |
BFS Ant 10 F (D) |
41.57 |
0.79 |
45.98 |
2.69 |
< .0001 |
BFS Post 10 F (mm) |
6.68 |
0.15 |
5.84 |
0.37 |
< .0001 |
BFS Post 10 F (D) |
50.55 |
1.12 |
58.01 |
3.78 |
< .0001 |
Elevation Ant 10 F (mm) |
0.007 |
0.004 |
0.085 |
0.122 |
< .0001 |
Elevation Post 10 F (mm) |
0.027 |
0.006 |
0.137 |
0.048 |
< .0001 |
Astigmatism (D) |
1.05 |
0.67 |
5.99 |
2.97 |
< .0001 |
Sim K max (D) |
43.06 |
0.99 |
56.18 |
6.44 |
< .0001 |
Sim K min (D) |
42.02 |
0.99 |
50.18 |
5.09 |
< .0001 |
Sim K mean (D) |
42.54 |
0.94 |
53.18 |
5.61 |
< .0001 |
Irregularity 3 mm |
1.05 |
0.43 |
6.53 |
2.63 |
< .0001 |
Irregularity 5 mm |
1.33 |
0.48 |
7.1 |
2.97 |
< .0001 |
Thinnest point thickness (microns) |
534.73 |
30.86 |
392.07 |
68.58 |
< .0001 |
Thinnest point location: X |
0.45 |
0.23 |
0.6 |
0.27 |
< .007 |
Thinnest point location: Y |
0.45 |
0.29 |
0.64 |
0.43 |
< .026 |
Difference of center-7mmthickness (microns) |
110.59 |
16.42 |
231.69 |
57.35 |
< .0001 |
INF/SUP difference of astigmatism (D) |
0.29 |
0.28 |
5.73 |
4.47 |
< .0001 |
BFS = Best Fitted Sphere; D = diopters; Ant = Anterior; Post = Posterior; 3T = centered on 3mm thinnest point; INF/SUP = Inferior/Superior, F=Float. |
Table 1: Orb scans indices in keratoconus and normal eyes.
|
BFS 10mm float |
BFS 3mm centered on thinnest point |
P value |
Anterior BFS measurement (D) |
41.57 |
42.49 |
< .0001 |
Posterior BFS measurement(D) |
50.55 |
54.56 |
< .0001 |
Anterior elevation (mm) |
0.007 |
0.001 |
< .0001 |
Posterior elevation (mm) |
0.027 |
0.003 |
< .0001 |
|
BFS 10mm float |
BFS 3mm centered on thinnest point |
P value |
Anterior BFS measurement(D) |
45.98 |
55.32 |
< .0001 |
Posterior BFS measurement(D) |
58.01 |
80.34 |
< .0001 |
Anterior elevation (mm) |
0.085 |
0.005 |
< .0001 |
Posterior elevation (mm) |
0.137 |
0.010 |
< .0001 |
Table 3: Comparison of the BFS 10mm float and BFS 3mm centered on thinnest point
measurements in KCN eyes. BFS: Best fitted sphere, D: diopters.
|
Area under the curve |
Cutoff points |
Anterior BFS 10mm float (D) |
98% |
42-43 |
Posterior BFS 10mm float (D) |
99.5% |
52-54 |
Anterior BFS 3mm centered on thinnest point (D) |
100% |
45.4 |
Posterior BFS 3mm centered on thinnest point (D) |
100% |
62.75 |
Anterior elevation on BFS 3mm centered on thinnest point (mm) |
86% |
0.0005-0.0155 |
Posterior elevation on BFS 3mm centered on thinnest point(mm) |
93% |
0.0035-8.44 |
Anterior elevation on BFS 10mm float (mm) |
100% |
0.022 |
Posterior elevation on BFS 10mm float (mm) |
100% |
0.052 |
- Parker JS, Van Dijk K, Melles GR (2015) Treatment options for advanced keratoconus: A review. Surv Ophthalmol 60: 459-480.
- Galvis V, Sherwin T, Tello A, Merayo J, Barrera R, et al. (2015) Keratoconus: an inflammatory disorder? Eye (Lond) 29: 843-859.
- Kymionis GD, Grentzelos MA, Portaliou DM, Kankariya VP, Randleman JB (2014) Corneal collagen cross linking (CXL) combined with refractive procedures for the treatment of corneal ectatic disorders: CXL plus. J Refract Surg 30: 566-576.
- Auffarth GU, Wang L, Volcker HE (2000) Keratoconus evaluation using the Orbscan Topography System. J Cataract Refract Surg 26: 222-228.
- Lim L, Wei RH, Chan WK, Tan DT (2007) Evaluation of keratoconus in Asians: role of Orb scan II and Tomey TMS-2 corneal topography. Am J Ophthalmol 143: 390-400.
- Rao SN, Raviv T, Majmudar PA, Epstein RJ (2002) Role of Orb scan II in screening keratoconus suspects before refractive corneal surgery. Ophthalmology 109: 1642-1646.
- Sonmez B, Doan MP, Hamilton DR (2007) Identification of scanning slit-beam topographic parameters important in distinguishing normal from keratoconic corneal morphologic features. Am J Ophthalmol 143: 401-408.
- Agarwal A, Jacob S, Agarwal S (2005) Posterior corneal changes in refractive surgery. In: Agarwal S, Agarwal A, Agarwal A, (ed)Step by step corneal topography.
- Fam HB, Lim KL (2006) Corneal elevation indices in normal and keratoconic eyes. J Cataract Refract Sur 32: 1281-1287.
- Tummanapalli SS, Potluri H, Vaddavalli PK, Sangwan VS (2015) Efficacy of axial and tangential corneal topography maps in detecting subclinicalkeratoconus. J Cataract Refract Surg 41: 2205-2214.
- Chan C, Ang M, Saad A, Chua D, Mejia M, et al. (2015) Validation of an Objective Scoring System for FormeFruste Keratoconus Detection and Post-LASIK Ectasia Risk Assessment in Asian Eyes. Cornea 34: 996-1004.
- Guilbert E, Saad A, Elluard M, Grise-Dulac A, Rouger H, et al. (2016) Repeatability of Keratometry Measurements Obtained With Three Topographers in Keratoconic and Normal Corneas. J Refract Surg 32: 187-192.
- Belin MW, Khachikian SS, Ambrosio JR, Salomao M (2007) Ectasia Detection with the Oculus Pentacam: Belin/ Ambrosio Enhanced Ectasia Display. Highlights of Ophthalmology 35: 5-12.
- Schlegel Z, Hoang-Xuan T, Gatinel D (2008) Comparison of and correlation between anterior and posterior corneal elevaion maps in normal eyes and keratoconus-suspect eyes. J Cataract Refract Surg 34: 789-795.
- Yazıcı AT, Pekel G, Bozkurt E, Yıldırım Y, Pekel E, et al. (2013) Measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients. Int J Ophthalmol 6: 521-525.
- Reinstein DZ, Archer TJ, Urs R, Gobbe M, RoyChoudhury A, et al. (2015) Detection of Keratoconus in Clinically and Algorithmically Topographically Normal Fellow Eyes Using Epithelial Thickness Analysis. J Refract Surg 31: 736-744.
- Prakash DP, Agarwal A, Agarwal A (2005) Detection of Subclinical Keratoconus by Orb scan. AIOC.
© by the Authors & Gavin Publishers. This is an Open Access Journal Article Published Under Attribution-Share Alike CC BY-SA: Creative Commons Attribution-Share Alike 4.0 International License. Read More About Open Access Policy.