A Novel and Sensitive LC/MS/MS Method for Quantitation of Pyochelin in Human Sputum Samples from Cystic Fibrosis Patients
Allena
J. Ji 1*, Dorothea Rat2, Stephane Muccio3,
Alexandra Tavernier3, Lynn Taylor1, Lee Lee Chung1,
Susan Richards1
1Biomarkers
and Clinical Bioanalyses-Boston, Sanofi, 1 the Mountain Road, Framingham, MA
01701, USA
2Biomarkers
and Clinical Bioanalyses-Frankfurt, Sanofi, Industriepark Hoechst, D-65926
Frankfurt am Main, Germany
3Biomarkers
and Clinical Bioanalyses-Montpellier, Sanofi, 371, Rue du Professeur Joseph
Blayac, 34184 Montpellier, France
*Corresponding
author: Allena J. Ji, Biomarkers and
Clinical Bioanalyses-Boston, Sanofi, 1 the Mountain Road, Framingham, MA 01701,
USA. Tel: 508-270-2519;E-mail: allena.ji@sanofi.com
Received Date: 31
May, 2019; Accepted Date: 20 June, 2019; Published
Date: 28 June, 2019
Pyoverdines and
pyochelin are siderophore compounds secreted by Pseudomonas
aeruginosa present in cystic fibrosis affected patients. The
available literature on the quantification of pyoverdines and
pyochelin in sputa of cystic fibrosis patients
is based on detecting the fluorescence properties of the targets
detecting pyoverdines and
pyochelin without differentiating the types by chromagraphic separation. Within
these studies pyochelin was
always far less detected than pyoverdines. This
is in contrast to gene expression data and suggets the
pyochelin fluorescence detection method lacks sensitivity which results in the
molecule being non-detectable. In this report, we
describe a novel and sensitive LC/MS/MS method for quantitation of pyochelin in
human sputum from cystic fibrosis affected patients with Pseudomonas
aeruginosa (culture positive) infection. Using a validated
bioanalytical method, 26 sputum samples from cystic fibrosis patients from a
biobank were analyzed. Data showed the LC/MS/MS method met bioanalytical
validation acceptance criteria and results of pyochelin were consistent with
reported results of Pseudomonas aeruginosa concentration in
cystic fibrosis patients. This new approach for quantitation of pyochelin
in human sputum is more sensitive, reproducible and easier to reliably measure
the biomarker pyochelin in cystic fibrosis patients. Monitoring pyoverdines and
pyochelin in the sputum samples from cystic fibrosis patients may provide
better analytical tools for cystic fibrosis new drug development.
1. Introduction
Cystic fibrosis is an
inherited autosomal recessive disease which is caused by the presence of
mutations in both copies of the gene for the cystic fibrosis transmembrane
conductance regulator protein [1]. The lungs of patients with cystic fibrosis
usually become chronically infected with Pseudomonas aeruginosa, a
bacterium that can acquire iron by secreting two kinds of iron-chelating
compounds (siderophores): pyoverdines and pyochelin (Figure 1).
Siderophore-regulated iron uptake is generally considered a key factor in
infections caused by Pseudomonas aeruginosa. The morbidity and
mortality associated with cystic fibrosis is mainly related to rapid lung
function decline and termed pulmonary exacerbation. There are considerable
evidences that Pseudomonas aeruginosa is present in the lungs
of the majority of cystic fibrosis patients, often in large amount i.e.>108 cfu/mL
sputum [2].
Pyverdine group has
more than 68 different peptide forms which has been characterized [3] and has
been classified as three subtypes, Type I, Type II, and Type III [4]. Each
subtype of pyverdine produced by individual strains is responsible for
distinctive yellow-green fluorescence of certain pseudomonads and has a
specific receptor for its uptake (FpvAI, FpvAII or FpvAIII, respectively) [4].
All subtypes of pyverdine have a strong ability to bind ferric ion with a
formation constant of 1022 to 1035 [5] or 1024 to
10 27 at pH 7.0 [6]. In mammals, most iron is bound to
proteins (primarily ferritin, transferrin, lactoferrin and hemoglobin) with
high affinity (Kd ~ 1020) to reduce iron bioavailability to
infecting bacteria. This action is an innate immunity against bacterial
infection. Pyverdines are primary siderophores secreted by Pseudomonosa
aeruginosa from infected cystic fibrosis patients to compete ferric
ions from normal iron holding protein like transferrin to support the bacteria
growth.
Pyochelin is
considered a secondary siderophore in Pseudomonosa aeruginosa, having
a much lower iron binding affinity than pyverdines at releasing iron from
transferrin [7-9]. It has been a major focus of research on Iron acquisition
by Pseudomonas aeruginosa related to pyverdine and pyochelin
[9-13]. The analytical methods for quantitation of pyverdine and pyochelin in
human sputum were based on fluorescence quenching with ferric chloride to form
ferripyoverdine, yellow-greenish color at excitation 378 nm and emission 390 to
550 nm [14] or at excitation 405 nm and emission over a range of or 425-530 nm
[15] for pyoverdine and at excitation 350 nm and emission 370 to 500 nm for
pyochelin. These methods do not have enough sensitivity; especially pyochelin
cannot be detected in cystic fibrosis patient sputa with Pseudomonas
aeruginosa positive or low detectable concentration of [15]. There is
not a pyochelin LC/MS/MS quantitation method for this type of application found
in published papers. In this article, we reported a new approach to quantify
pyochelin using LC/MS/MS. The method was developed and validated in a linear
range of 0.25 to 25µM in human sputum. We also analyzed pyochelin concentration
from 26 sputa with known Pseudomonas
aeruginosa values from cystic
fibrosis pateints and found detectable
pyochelin concentrations in some of these samples.
2. Materials
and Methods
2.1. Reagents
and equipment
Pyochelin was
synthesized by Sanofi R&D, Parallel Synthesis & Natural Products
Chemistry, Vitry, Lot VAC.XFQ8.225.2, Molecular weight 324.42, purity
69.2%. Internal standard (IS), 13C4-15N-pyochelin
(product code RA11654539, molecular weight 329.37) was synthesized by Isotope
Chemistry and Metabolite Synthesis, Sanofi, Frankfurt, Germany. Acetonitrile
(HPLC grade), and methanol (HPLC grade) were purchased from Burdick &
Jackson (distributed by VWR Scientific Products, Newark, NJ, USA).
Ammonium acetate, PBS
Buffer (10× concentrate, pH 7.4), Tween® 20, Bovine Serum Albumin (BSA), and
DL-Dithiothreitol (DTT) Solution (~ 1M in H2O) were purchased from
Sigma Aldrich (St. Louis, MO, USA). Deionized water was obtained from an
in-house Milli-Q DI System (Millipore, Billerica, MA, USA).
Human sputum samples
collected from donors without cystic fibrosis were purchased from
Bioreclamation, Westbury, New York, USA. Human sputum samples from cystic
fibrosis patients with for Pseudomonas aeruginosa results were
purchased from BioPartner, Woodland Hills, CA, USA.
The triple quadrupole
mass spectrometer, model Sciex API 4000 was manufactured by Applied
Biosystems (Toronto, Canada). Acquity UPLC System (H class) was manufactured by
Waters Corporation (Milford, MA, USA). Multi-tube vortex was from VWR
Scientific Products (Bridgeport, NJ). An HPLC column, Chromenta™ KB-SiO2 HILIC,
2.1 mm × 100 mm, 120Å, 3 µm was purchased from Columnex LLC, San Diego, CA,
USA. Centrifuge, AllegraTM 6R was manufactured by Beckman
Coulter Inc. Indianapolis, IN, USA; Eppendorf micro centrifuge was made by
Brinkman, Boston, MA, USA; MaxQ™ 4450 benchtop incubating orbital shaker was
made by Thermo Scientific; and Sonicator was manufactured by Thomas Scientific
Inc., Swedesboro, NJ, USA.
2.2. Preparation
of homogenized sputum
For the matrix of
pyochelin standards and quality control samples, individual human sputum
samples (0.5 to 4 g, 15-20 lots from non-CF patients with Pseudomonas
aeruginosa results) were transferred to pre-weighted 15-mL conical
polypropylene tubes with cap and weighed, respectively. The net sputum weight
in each tube was calculated. Using 1:1 (w/v) ratio, a volume of homogenizing
solution (PBS Buffer (1×) with 0.1% (v/v) Tween® 20, 2% (w/v) BSA, 0.2% DTT)
was added to each sputum sample tube to thin the viscosity of sputum. The
sample tubes were vigorously vortexed and then incubated on an orbital shaker
at 400 rpm at Room Temperature (RT) for 45 minutes. Then the sample tubes were
vortexed briefly and centrifuged at 1000 × g for 30 minutes at RT. The top
clear homogenized sputum solution (homogenized sputum) from multiple sample
tubes (control sputa without detectable pyochelin) were combined, mixed into
one 50-ml centrifuge tube, then the pooled homogenized sputum solution was
aliquoted to 4.0 mL per tube with screw cap and stored at -80ºC for
future standards and QC preparation and extraction. For individual cystic
fibrosis patient sputum samples, the homogenized sputum samples are prepared
according to the above procedure individually (not pooled) and the homogenized
sputum from each patient sample was stored at -80 ºC for
further extraction.
2.3. Preparation
of pyochelin standards and QCs in homogenized sputum
Primary pyochelin
stock solution, 2.50 mM, (0.811 mg/mL, molecular weight 324.42, with correction
of purity 69.2%) was prepared in methanol. The two independently weighted
pyochelin stock solutions were prepared, diluted, and chromatographically
compared to be ≤ 10% difference to confirm the correctness. The two stock
solutions (Stock A for calibration standards and Stock B for QCs) were then
aliquoted to 100 µL per 2-ml HPLC vial for multiple vials. These HPLC vials
were evaporated under nitrogen flow at 37 ºC heat block to dryness. All dry
vials were stored at -20ºC for future analysis use. At the time of sample
analysis, a dry vial of pyochelin was reconstituted with 100 µL of methanol,
vortexed, sonicated for 5 minutes at RT; then 900 µL of methanol was added to
the same vial to make 250 µM pyochelin secondary solutions. Calibration
Standards (400 -1000 µL) were prepared by mixing appropriate volume of 250 µM
pyochelin stock solution A with the blank homogenized sputum to generate eight
concentrations at 0.125, 0.250, 0.500, 1.25, 2.50, 5.00, 11.25, and 12.5µM
which were corresponding to final sputum pyochelin concentrations of 0.250,
0.500, 1.00, 2.50, 5.00, 10.0, 22.5, and 25.0 µM due to 1:1 w/v ratio mix at
homogenizing step. Quality Control (QC) samples (600 - 1500 µL) were prepared
in a similar way as standards using pyochelin 250 µM stock solution B mixed
with the blank sputum homogenizing supernatant to generate LLOQ,
low, mid, and high QC at 0.125, 0.375, 3.75, and 10.0 µM, which were
corresponding to final sputum pyochelin concentrations of 0.250, 0.750, 7.50,
and 20.0 µM. Both calibration standards and QCs were prepared freshly for each
run in the method validation. A system validation solution containing 0.00852
µM Pyochelin and 0.0710 µM IS in solvent of 80% Acetonitrile, 20% Water, 0.01%
DTT (v/v/v) was prepared for daily system suitability check.
2.4. Extraction
procedure for homogenized sputum
Aliquots of 100 µL of
homogenized sputum calibration standards, QC samples, blank homogenized sputum,
and unknown cystic fibrosis patient homogenized sputum samples were pipetted to
pre-labelled 1.5-mL micro-centrifuge tubes. Aliquots of 1.0 mL of internal
standard working solution (0.312 µM 13C4-15N-pyochelin
in 100% acetonitrile) were added to each tube except the samples without
internal standard (1.0 mL of IS solvent, acetonitrile used for this sample).
All samples were vortexed in a multi-vortex at 2000 rpm for one minute for
twice and then centrifuged in Eppendorf centrifuge at 13200 rpm for 5 minutes.
The supernatant 100 µL from each tube was transferred to a 2-mL clean HPLC
glass vial, then diluted with 300 µL of the diluent (80% Acetonitrile, 20%
Water, 0.01% (v/v) DTT), and mixed briefly. The diluted supernatant samples (10
µL) were submitted for LC/MS/MS analysis.
2.5. LC/MS/MS
Conditions
Liquid chromatographic
separation procedures were carried out using a Chromenta™ KB-SiO2 HILIC
column, 2.1 mm × 100 mm, 120Å, 3 µm (Columnex LLC) at room temperature with a 5
minutes’ gradient program at the flow rate of 0.4 mL/min on a Waters Acquity
UPLC system with mobile phase A (MPA, 50% Acetonitrile, 50% Water, 10 mM
Ammonium Acetate) and mobile phase B (MPB, 95% Acetonitrile, 5% Water, 10 mM
Ammonium Acetate). The gradient program was at 0 to 3.5 minutes with a ramp 5%
MPA to 50% MPA, 3.5 to 3.6 minutes from 50% to 5% MPA, and 3.5 to 5.0 minutes
stayed 5% MPA. The UPLC auto sampler temperature was maintained at 4 ºC.
The triple quadrupole Sciex API-4000 mass spectrometer was operated at positive
electrospray ionization mode in multiple reaction monitoring (MRM) mode, Ion
source temperature at 550 °C, dwell time 150 milliseconds. The mass
spectrometry transitions were: m/z 325.6→191.3 for
Pyochelin and 330.2 →151.1 for IS. Both m/z transitions were
using decluttering potential 70 volt (v), entrance potential 10v, and collision
energy 35v. The collision cell exit potential was 11.0v for Pyochelin and 8.0v
for IS.
2.6. Pyochelin
method validation and sample analysis
According to US Food
and Drug Administration bioanalytical method validation for industry guidance
May 2018 [16,17], the method validation elements of specificity, sensitivity,
linearity, precision and accuracy, dilution integrity, stability, carryover,
extraction recovery, and matrix effect were evaluated during method
validation. The software, analyst (version 1.6.1, AB Sciex, Toronto,
Canada), was used for MS/MS data processing. Peak area ratios of pyochelin to
internal standard were plotted versus a calibration curve. Linear regression
model employing a 1/x weighting was used to calculate the concentrations in QC
and unknown samples. There was not endogenous pyochelin found in pooled
homogenized sputum matrix. Twenty-six sputum samples from
BioPartners with pre-measured Pseudomonas aeruginosa values
were individually homogenized with the homogenization buffer described in this
article. The supernatant solutions of these homogenized sputum lots were
extracted and analyzed on the LC/MS/MS system.
3. Results
and Discussion
3.1. Specificity
and sensitivity
The ion chromatogram
of pyochelin and its internal standard from a non-CF patient sputum blank 00,
a lower limit of quantitation (LLOQ)
0.25 µM pyochelin, and a CF patient sputum sample acquired in the API-4000
conditions described in the experimental section are presented in
(Figure 2). The chromatograms indicated adequate signal to noise at LLOQ,
being much higher than 10 fold. The method showed good specificity for
pyochelin quantitation. The detection limit was 0.011 µM (3x of noise). The
LLOQ of human plasma was validated 0.250 µM. The intra-and inter-day precision
was expressed in terms of the coefficients of variation within a run and among
runs using spiked LLOQ and QC samples. Intra-day precision of the LLOQ was from
5.0 to 10.9% for CV and intra-day accuracy (%Bias) was from 2.6 to 5.8% except
the 2nd run, 21.1%. Inter-day precision and accuracy for all 4
runs were 10.4% for CV and 8.2% for bias (n =20), respectively (Table
1, the first column). The sensitivity of the assay passed validation
acceptance criteria. In our later analytical work, the API-4000 mass
spectrometer was changed to Qtrap-6500 so that the assay sensitivity was able
to reach to 0.030 µM which was eight time sensitive than the reported method.
3.2. Linearity
The pyochelin human
sputum standard curve was validated from 0.25 to 25.0 µM using approximately
0.5 to 10 g of human sputum. Pyochelin concentrations (µM) were obtained using
1/x2 weighted linear regression analysis of the 8 calibration
standards after comparison of mean absolute bias using weighted linear
regression at 1, 1/x and 1/x2, respectively. All standard
curves from the four core validation runs had coefficients of determination
(r2) ≥ 0.9958. During method development, we found pyochelin calibration curve
was not linear.
The concentration of
DDT in final injection solution was optimized and determined that only when DDT
concentration was in range of 0.01 to 0.05%, it could produce a constant
response factor of pyochelin. Therefore, 0.01% DDT(v/v) was used in 300 µL of
the diluent (80% Acetonitrile, 20% Water, 0.01% (v/v) DTT) for final
supernatant dilution before LC/MS/MS analysis.
3.3. Precision
and accuracy
The
intra-day Accuracy (% Bias) QC samples at three different concentrations
(0.750, 7.50 and 20.0 µM) ranged from -8.5 to 7.7% with precision (CV) ranging
from 1.8 to 12.9%. Inter-day accuracy ranged from 2.8 to 6.9% for bias with a
CV range from 8.0 to 11.6% (Table 1, the last 3 columns). It
was noted that the second core validation run had a bias of 21.1% for LLOQ,
21.3% for QCL, 18.1% for QCM, and 18.0% for QCL. This run should not be
rejected because there was not an identified cause in operation. Our validation
acceptance criteria were set as (a) at least 3 inter-day runs must pass
intra-day acceptance criteria; (b) if any intra-day runs fail to meet
acceptance criteria, the last two runs must consecutively pass the intra-day
acceptance criteria; and (c) overall inter-day precision and accuracy for all
runs must pass inter-day acceptance criteria. Therefore, overall the validation
performance passed intra-day and inter-day acceptance criteria even if the 2nd run
failed to meet %bias acceptance criteria.
It
is worth noting that due to mixing of 1:1 weight to volume ratio of patient
sputum sample to the homogenizing solution, the actual pyochelin concentration
in the final homogenized sputum is only half of the patient sputum
concentration. Carryover impact was performed by injecting an extracted sputum
blank 00 (no Pyochelin or IS) after the highest standard (25.0 µM) in all
validation runs that contained a standard curve. Carryover impact to LLOQ was
calculated using the analyte or IS peak area counts in blank 00 after standard
8 divided by the analyte or IS peak area counts in Standard 1, respectively.
The carryover from 7 validation runs was ≤ 7.1% for pyochelin and 0% for IS.
Therefore, the carryover was acceptable for both pyochelin (≤ 20% of LLOQ) and
IS (≤ 5.0% of added IS) and does not have an impact on the analytical method.
3.4. Post-extract
dilution integrity
Post-extract dilution
integrity was performed at dilution quality control samples (QCD2X and QCD5X,
40 µM) using the concentration of two times (2X) of the QCH level. The QCD2X
and QCD5X samples were first extracted in 5 replicates per group without matrix
dilution.
The final injection
solution was diluted with a pool of extracted blank 0 (blank with IS) in volume
of 1:1 or 1:4 of each extracted QCD2X or QCD5X sample to extracted sputum blank
0. The dilution integrity results showed that post-extract dilution had %CV
from 4.0 to 5.0 % and bias ranged from -5.4 to -2.0%,
which met acceptance criteria (Table 2). This type of dilution is more
convenient than the matrix dilution before sample extraction and saves time for
repeat sample analysis. Sputum blank with IS (control sputum) from the same run
was used as the diluent for post-extraction dilution which kept the IS
concentration similar to the IS counts from standards after dilution but only
the analyte concentration was diluted. In this way, pyochelin concentration in
the diluted sample could be calculated from calibration curve and then
multiplied by the dilution factor.
3.5. Stability
Stability of pyochelin
was evaluated during method validation in various conditions. Results (Table 3)
demonstrated that pyochelin was stable after three cycles of freeze–thaw at ‑80ºC/ RT and
stable for up to 17 days at -80 ºC in homogenized sputum
samples. Pyochelin was stable for up to 72 hours at 4 ºC after
reinjecting extracted samples using the initial injected standard curve at time
zero (or baseline) for its concentration regression. Pyochelin was also stable
and reproducible after 72 hours at 4 ºC and then reinjecting
the entire run using reinjected standard curve to regress the concentration
(comparing to time zero results from the regression of baseline standard
curve). Pyochelin was also stable for up to 34 days at -20ºC in
methanol (stock solution) and 24 days at -20 ºC in 80%
acetonitrile/20% water/0.01% DTT solution in the system validation solution.
3.6. Extraction
recovery
Extraction recovery
was assessed by comparing the mean peak area of extracted (QCL, QCM and QCH) at
each group to the mean peak area of extracted matrix blank without IS that was
post-spiked with Pyochelin and IS in each group. Pyochelin or IS extraction
recovery was evaluated at three concentration levels (QCL, QCM and QCH), each
in five replicates (n = 5), respectively. The mean % extraction recovery, SD,
and % CV at n = 5 were calculated using group mean of the peak area from
extracted sample divided by mean of peak area from the post-spiked samples for
pyochelin or IS, respectively. The overall mean of extraction recoveries for
the analyte and IS are calculated from the three concentration group means and
%CV was calculated also. The results in Table 4 show that the mean extraction
recoveries were 99.2% for Pyochelin and 96.5% for the IS. The overall %CV from
the three pyochelin concentration groups was ≤ 7.4% which met acceptance
criteria ≤ 15%. There was not an analyte concentration dependent issue on the
analyte extraction recovery. The data showed near 100% extraction recoveries
for Pyochelin and IS in the sputum assay.
It was observed %CV
> 20% in two groups of peak area for Pyochelin and IS in QCH pre-spiked
extraction group, it should not impact on pyochelin quantitation due to peak
area ratio used for calculation of the analyte concentration. The stable
labelled 13C4-15N-pyochelin is used as
internal standard for this method, which is very important to achieve the
sputum mass spectrometry assay. An analog IS
(2-(4-Fluorophenyl)-4-methyl-1,3-thaiazole-5-carboxylic acid, molecular weight
237.25) was used during the method development for sample extraction in early
stage of method development when stable labelled internal standard was not
available. The analog internal standard showed non-parallel extraction efficiency
with Pyochelin.
Sputum is not a
typical secreted fluid from normal subjects’ respiratory system, therefore, it
can be collected from respiratory disease patients only. Also, sputum is not a
homogeneous liquid so that it makes analysis of pyochelin difficult. The key
procedure of the extraction method was to homogenize heterogeneous human sputum
to a more uniform liquid. We developed 1:1 (w/v) ratio of sputum weight to a
volume of homogenizing solution (PBS Buffer (1×) with 0.1% (v/v) Tween® 20, 2%
(w/v) BSA, 0.2% DTT) to reduce the viscosity of sputum first, then vigorously
vortexed and incubated samples at RT on an orbital shaker for 45 minutes to
make a homogenized sputum samples. The 0.2% DDT in homogenizing solution was
for breaking down thiol bonds between pyochelin or pyoverdine and various
proteins in the sputum.
The 0.1%Tween 20 was
used for preventing adsorption of pyoverdine from the container walls which was
necessary for the ELISA assay and incorporated because the same homogenized
sputum sample was used for multiple bioanalyses and shared by different sites.
Fortunately, due to the large dilution in our LC/MS/MS assay extraction, the
tween 20 (the suppression effect for mass spectrometer ionization) did not
impact the LC/MS/MS quantitation of the analyte. The 2% BSA was for a better
homogenization of sputum to a solution. After the sample tubes were vortexed
and centrifuged, the supernatant of the homogenized sputum sample was stored at
‑80 ºC for further protein precipitation extraction and
LC/MS/MS analysis. During method development, we found that it was unsuccessful
to use artificial sputum to prepare standard curve, because the extracted
artificial sputum had a strong chromatographic interference and high viscosity
so that the analytical column pressure was too high to operate. The reported
analytical homogenization and extraction methods worked well for pyochelin
quantitation without sacrificing LC pressure condition.
Absolute matrix effect
was assessed by comparing the peak area counts of post-extracted blank sputum
to neat solution. Both post-extracted sputum and neat solution samples (100µL)
were spiked with Pyochelin and IS (300 µL of 0.0114 µM pyochelin and 0.0946 µM
IS for QCL, 0.114 µM pyochelin and 0.0946 µM IS for QCM, and 0.303 µM pyochelin
and 0.0946 µM IS for QCH) to reach theoretical concentration of extracted QCL,
QCM or QCH in the solvent of 80% Acetonitrile, 20% Water, 0.01% DTT. The peak
area counts of Pyochelin and IS in the corresponding post-extracted sputum and
neat solutions were evaluated for matrix effect. Absolute Matrix Effect (%) is
calculated by subtracting mean neat solution peak area from the post-spiked QC
peak area, then dividing by mean neat solution peak area, and finally
multiplying 100. Table 5 showed that matrix effect was -5.6% for Pyochelin and
-5.9 % for IS. The results indicate a minor ion suppression for both pyochelin
and IS during mass spectrometry analysis. However, the level of suppression for
Pyochelin and its stable labelled IS was consistent. Use of peak area ratio for
quantitation compensates this suppression effect. The relative matrix effect
assessment across multiple sputa was not evaluated in this method validation
due to the rare matrix of sputa from cystic fibrosis patients.
3.8. Clinical
application
Using this reported
analytical method, the concentrations of pyochelin from 26 cystic fibrosis
patient samples were analyzed. The results are listed in Table 6. All 26 cystic
fibrosis patient sputa from BioPartners Inc. were cultured for Pseudomonas
aeruginosa and other microbiology organism bacteria counts by
the vender before the samples were received at our site. These bacteria counts
are also reported with pyochelin results. There were 9 out of 26 sputum samples
with detectable Pyochelin concentrations, among the 9 sputum lots, 5 sputum
lots were above LLOQ using the validated method. The chromatograms of
quantified pyochelin and detected pyochelin are showed in Figure 2. Our method
was subsequently transferred to another site and partially validated using Qtrap-6500
mass spectrometer. The assay LLOQ was lowered to 30 nM which was 8.3 folds
lower than reported LLOQ (0.250 µM) in this article. Using the more sensitive
mass spectrometer, the pyochelin concentration was quantified in sputa from
cystic fibrosis patients (additional set of cystic fibrosis patient samples).
These results were consistent with pyoverdine concentrations tested in the same
sputa with Pseudomonas aeruginosa culture positive results
(data not shown here).
4. Conclusions
A novel and sensitive analytical
approach for quantitation of pyochelin in sputa from cystic fibrosis patients
was developed and validated using LC/MS/MS. The method was demonstrated to be
accurate and reproducible for use in a clinical setting. The performance
characteristics of the assay met the bioanalytical method acceptance criteria.
Using the reported method, the concentrations of pyochelin from cystic fibrosis
patients were analysed. Quantifiable pyochelin concentrations in these cystic
fibrosis patient sputa also had high corresponding Pseudomonas
aeruginosa culture results.
However, given the method sensitivity, high positive Pseudomonas
aeruginosa results are not a prerequisite to
find detectable pyochelin. Per our knowledge, this is the first
mass spectrometry quantitation method for pyochelin in human sputum. This
LC/MS/MS assay provides a new and sensitive approach to quantify pyochelin in
human sputum and its use as a biomarker may enhance new drug development for
cystic fibrosis.
6. Acknowledgement
Authors would like to
thank Haixing Wang for her participation of method development (left company),
Damen Sallabery for synthesis of pyochelin reference standard and 13C-phyochelin
internal standard, and Patrick Decourcy for his effort to obtain patient sputa
with Pseudomonas aeruginosa culture results.
6. Conflicts
of Interest
Figure
1: Chemical Structure of pyochelin and its
internal Standard.
Figure 2: The ion chromatograms of pyochelin from extracted human sputum samples. (a) blank human sputum from a non-cystic fibrosis patient; (b) Spiked Pyochelin at LLOQ 0.25 µM in the blank sputum; (c) a cystic fibrosis patient sputum with pyochelin concentration at 1.33 µM.
|
QC ID Conc.,
µM |
LLOQ 0.250 |
QCL 0.750 |
QCM 7.50 |
QCH 20.0 |
Intra-Run 4 Runs |
% CV |
5.0 to 10.9 |
2.8 to 4.7 |
2.5 to 12.9 |
1.8 to 5.1 |
% Bias |
2.6 to 5.8a |
-8.5 to -0.7b |
-4.0 to 7.7b |
-0.3 to 1.3b |
|
n/Run |
5 |
5 |
5 |
5 |
|
Global Inter-Run |
% CV |
10.4 |
11.6 |
10.2 |
8.0 |
% Bias |
8.2 |
2.8 |
6.9 |
5.0 |
|
n |
20 |
20 |
20 |
20 |
|
a)
LLOQ of
Runs 1, 3 and 4’s intra-day and overall 4 runs inter-day met ≤ ± 20% bias
acceptance criteria except the 2nd run, LLOQ’s bias was 21.0%. b)
Precision
and accuracy of QCL, QCM, and QCH in Runs 1, 3 and 4’s intra-day and overall
4 run inter-day met ≤ ±15% acceptance criteria except the 2nd run,
the bias was 21.3% for QCL, 18.1% for QCM, and 18.0% for QCL. |
Table
1: Precision and Accuracy of Pyochelin in Human
Sputum.
Dilution Factor 2X |
Dilution Factor 5X |
|
Mean |
39.2 |
37.8 |
SD |
1.55 |
1.88 |
%CV |
4.0 |
5.0 |
%Bias |
-2.0 |
-5.4 |
n |
5 |
5 |
The detailed
post-extraction dilution procedures: 2X = 100 µL final
sample solution mix with 100 µL of extracted blank 0
solution (with IS) and inject the sample. 5X = 100 µL final
sample solution mixed with 400 µL of extracted blank 0 solution (with IS) and
inject the sample. |
Table
2: Dilution Integrity of Pyochelin in the Post Extracted
Sample.
Name |
Matrix |
Conditions |
Results |
Freeze/Thaw
Stability |
Homogenized Human
Sputum |
3 cycles at -80 ºC/RT |
%CV ≤ 6.3%, %Bias ≤
± 10.7% |
Long Term Stability |
Homogenized Human
Sputum |
17 days at -80 ºC |
%CV ≤ 5.5%, %Bias ≤
± 12.3% |
Re-injection
Stability |
Diluted extracted
sputum solution (82.5%ACN/ 17.5%Water/0.0075%DTT) |
72 hours at 4 ºC |
%CV ≤ 3.8%, %Bias ≤
± 5.1% |
Reinjection
Reproducibility |
Diluted extracted
sputum solution (82.5%ACN/ 17.5%Water/0.0075%DTT) |
72 hours at 4 ºC |
%CV ≤ 3.7%, %Bias ≤
± 6.5% |
Stock Solution
Stability |
Methanol |
34 days at -20 ºC |
% Difference from
the fresh solution, 4.0% |
SVS Stability |
80%ACN/20%Water/0.01%DTT |
24 days at -20 ºC |
% Difference from
the fresh solution, 6.8% |
Table
3: Pyochelin Stability in Homogenized Human
Sputum, Processed Samples, and Solutions.
Validation Sample ID |
Statistics
Parameters |
|
Pyochelin |
|
|
IS |
|
Pre-Spiked Peak
Area |
Post-Spiked Peak
Area |
Analyte
Recovery% |
Pre-Spiked Peak
Area |
Post-Spiked Peak
Area |
IS Recovery% |
||
QCL 0.75 µM |
Mean |
3042 |
3065 |
99.2 |
27385 |
30286 |
90.4 |
S.D. |
236 |
172 |
|
2060 |
849 |
|
|
%CV |
7.8 |
5.6 |
|
7.5 |
2.8 |
|
|
n |
5 |
5 |
|
5 |
5 |
|
|
QCM 7.50 µM |
Mean |
26655 |
26206 |
100.9 |
25471 |
24430 |
104.3 |
S.D. |
2842 |
1383 |
|
3483 |
1431 |
|
|
%CV |
10.7 |
5.3 |
|
13.7 |
5.9 |
|
|
n |
5 |
5 |
|
5 |
5 |
|
|
QCH 20.0 µM |
Mean |
89056 |
91427 |
97.4 |
30971 |
32697 |
94.7 |
S.D. |
18052 |
5786 |
|
9465 |
2784 |
|
|
%CV |
20.3 |
6.3 |
|
30.6 |
8.5 |
|
|
n |
5 |
5 |
|
5 |
5 |
|
|
Overall Mean Recovery |
|
|
99.2 |
|
|
96.5 |
|
SD |
|
|
1.75 |
|
|
7.12 |
|
%CV |
|
|
1.8 |
|
|
7.4 |
|
n |
|
|
3 |
|
|
3 |
Table
4: Extraction Recovery of Pyochelin and IS in
Human Sputum.
Sample Name |
|
Pyochelin |
|
|
IS |
|
||
Post-spiked |
Neat Solution |
Mean Neat Solution |
% Matrix Effect |
Post-spiked |
Neat Solution |
Mean Neat Solution |
% Matrix Effect |
|
QCL |
10930 |
11690 |
11708 |
-6.6 |
80590 |
88650 |
86678 |
-7.0 |
11730 |
11860 |
0.2 |
84910 |
86860 |
-2.0 |
|||
10950 |
12020 |
-6.5 |
83260 |
87830 |
-3.9 |
|||
11100 |
12420 |
-5.2 |
79640 |
87190 |
-8.1 |
|||
11840 |
10550 |
1.1 |
83360 |
82860 |
-3.8 |
|||
QCM |
103000 |
109700 |
114200 |
-9.8 |
82350 |
86720 |
88896 |
-7.4 |
105900 |
118200 |
-7.3 |
83700 |
86310 |
-5.8 |
|||
106900 |
114200 |
-6.4 |
84040 |
88640 |
-5.5 |
|||
105000 |
111200 |
-8.1 |
81090 |
91120 |
-8.8 |
|||
107800 |
117700 |
-5.6 |
82070 |
91690 |
-7.7 |
|||
QCH |
289700 |
305200 |
312420 |
-7.3 |
88070 |
96000 |
95166 |
-7.5 |
297600 |
308300 |
-4.7 |
92500 |
92170 |
-2.8 |
|||
296400 |
306600 |
-5.1 |
88670 |
95160 |
-6.8 |
|||
290100 |
316700 |
-7.1 |
88660 |
95210 |
-6.8 |
|||
296400 |
325300 |
-5.1 |
90300 |
97290 |
-5.1 |
|||
%Overall Mean Matrix Effect |
-5.6 |
|
-5.9 |
Table
5: Matrix Effect of Pyochelin and IS in
Extracted Human Sputum.
Sample # |
Sample Name |
Conc. (µM) |
Comments |
1 |
SPUTUM_CF
PATIENT_0910 |
BQL |
NN - solitary;
EA 9×104cfu/ml; PA 5×10²cfu/ml; SA 5×10³cfu/ml; CA 1×104cfu/ml
|
2 |
SPUTUM_CF
PATIENT_0911 |
BQL |
NN - solitary;
SA 5×105cfu/ml |
3 |
SPUTUM_CF
PATIENT_0912 |
BQL |
PA 1×105cfu/ml;
SA 1×108cfu/ml |
4 |
SPUTUM_CF
PATIENT_0913 |
BQL |
NN - solitary;
SE - solitary; CA 5×10 cfu/ml |
5 |
SPUTUM_CF
PATIENT_0916 |
BQL |
SA 1×104cfu/ml;
BHS (not group A) 1×10³cfu/ml |
6 |
SPUTUM_CF
PATIENT_0918 |
BQL(0.110*) |
PA 1×108cfu/ml |
7 |
SPUTUM_CF
PATIENT_0919 |
BQL (0.209*) |
NN - solitary;
SVs - solitary; PA 5×10²cfu/ml; SA 1×10³cfu/ml; CA 5×10²cfu/ml |
8 |
SPUTUM_CF
PATIENT_0921 |
0.284 |
PA 1×108cfu/ml;
SA 5×104cfu/ml |
9 |
SPUTUM_CF
PATIENT_0922 |
0.324 |
PA 5×106cfu/ml;
SA 1×108cfu/ml |
10 |
SPUTUM_CF
PATIENT_0935 |
BQL |
PA 2×102cfu/ml;
SAs 1×104cfu/ml |
11 |
SPUTUM_CF
PATIENT_0936 |
BQL |
SAH 1×105
cfu/ml; CA 1×104 cfu/ml |
12 |
SPUTUM_CF
PATIENT_0939 |
BQL (0.171*) |
PA; SA |
13 |
SPUTUM_CF PATIENT_0941 |
BQL |
PA 1×108
cfu/ml; SA 1×106 cfu/ml |
14 |
SPUTUM_CF
PATIENT_0942 |
1.33 |
PA 1×108
cfu/ml; SA 1×108 cfu/ml |
15 |
SPUTUM_CF
PATIENT_0945 |
BQL(0.031*) |
PA 108
cfu/ml |
16 |
SPUTUM_CF
PATIENT_0946 |
BQL |
CA and other
Candida.nonalbicans 106 cfu/ml; SA 108 cfu/ml |
17 |
SPUTUM_CF
PATIENT_0947 |
BQL |
PA 1×106
cfu/ml; SA 108 cfu/ml |
18 |
SPUTUM_CF
PATIENT_0948 |
BQL |
PA 1×108cfu/ml;SA
1×108cfu/ml |
19 |
SPUTUM_CF
PATIENT_0954 |
0.390 |
PA1×106cfu/ml;
CA 5×10 cfu/ml |
20 |
SPUTUM_CF
PATIENT_0955 |
0.294 |
PA 1×105cfu/ml
|
21 |
SPUTUM_CF
PATIENT_0956 |
BQL |
PA 5×105cfu/ml;
CA 5×102cfu/ml;EF 5×103cfu/ml; |
22 |
SPUTUM_CF
PATIENT_0957 |
BQL |
PA 5×103cfu/ml;
CA 1×102cfu/ml; SA 1×103cfu/ml; EF 1×105cfu/ml
|
23 |
SPUTUM_CF
PATIENT_0960 |
BQL |
PA 5×105cfu/ml;
SA1×104cfu/ml |
24 |
SPUTUM_CF
PATIENT_0970 |
BQL |
SA 5×10³cfu/ml;
SV |
25 |
SPUTUM_CF
PATIENT_0971 |
BQL |
CA 1×10 cfu/ml;
EC 1×10³cfu/ml; SA 5×10³cfu/ml |
26 |
SPUTUM_CF
PATIENT_0973 |
BQL |
PA 5×103
cfu/ml; SA 5×103 cfu/ml |
Note:
*Concentration level is below quantitation limit (≤ 0.250 µM, BQL). |
|||
BHS: Beta-hemolytic streptococcus; CA: Candida
albicans; EC: Enterobacter cloacae ; EF: Enterococcus faecium; NN:
Nonpathogenic neisseria ; PA: Pseudomonas aeruginosa; SA: Staphylococcus
aureus; SAH: Streptococcus a-haemolyticus; SE: Staphylococcus
epidermidis; SV: Streptococcus viridans. |
Table 6: Pyochelin Concentration in Sputa of Cystic Fibrosis Patients.
- Davis PB (2006) Cystic fibrosis since 1938. Am J Respir Crit Care Med 173: 475-482.
- Reid DW, O’Mary C, Champion A, Kirov SM (2007) Increased airway iron as a potential factor in the persistence of Pseudomonas aeruginosa infection in cystic fibrosis. Eur Respir J 30: 286-292.
- Meyer JM, Gruffaz C, Raharinosy V, Bezverbnaya I, Schafer M, et al. (2008) Siderotyping of fluorescent Pseudomonas: molecular mass determination by mass spectrometry as a powerful pyoverdine siderotyping method. Biometals 21: 259-271.
- Meyer JM, Stintzi A, De Vos D, Cornelis P, Tappe R, et al. (1997) Use of siderophores to type pseudomonads: the three Pseudomonosa aeruginosa pyoverdine systems. Microbiology 143: 35-43.
- Matzanke BF (2005)
Iron Transport: Siderophores. In: King B (ed) Encyclopedia of Inorganic
Chemistry, 2nd edn. Wiley, New York.
- Budikewicz H (2004) Siderophores of pseudomonodaceaesensu stricto (fluoresencent and non-fluororescent Pseudomonoa spp.). Fortschr Chem Org Naturst 87: 81-237.
- Sriyosachati
S, Cox D (1986) Siderophore-mediated iron acquisition from transferrin by
Pseudomonas aeruginosa. Infect Immun 52: 885-891.
- Cornelis P (2010) Iron uptake and metabolism in pseudomonads. Appl Microbiol Biotechnol 86: 1637-1645.
- Lamont IL, Konigs AF, Reid DW (2009) Iron acquisition by Pseudomonas aeruginosa in the lungs of patients with cystic fibrosis. Biometals 22: 53-60.
- Reid DW, Anderson GJ, Lamont IL (2009) Role of lung iron in determining the bacterial and host struggle in cystic fibrosis. Am J Physiol Lung Cell Mol Physiol 297: L795-L802.
- Koning AF, Martin LW, Sharples KJ, Roddam LF, Latham R, et al. (2013) Pseudomonas aeruginosa uses multiple pathways to acquire Ion during chronic Infection in cystic fibrosis lungs. Infection and Immunity 81: 2697-2704.
- Braud A, Geoffroy V, Hoegy F, Mislin GLA, Schalk IJ (2010) Presence of the siderophores pyoverdine and pyocheln in the extracellular medium reduces toxic metal accumulation in Pseudomonas aeruginosa and increases bacterial metal tolerance. Env Microbiol Rep 2: 419-425.
- Saha R, Saha N, Donofrio RS, Bestervelt LL (2013) Microbial siderophores: a mini review. J Basic Microbiol 53: 303-317.
- Haas B, Kraut J, Marks J, Zanker SC, Castignetti D (1991) Siderophore presence in sputa of cystic fibrosis patients. Infect Immun 59: 3997-4000.
- Martin LW, Reid DW, Sharples KJ, Lamont IL (2011) Pseuodomonas siderophores in the sputum of patients with cystic fibrosis. Biometals 24: 1059-1067.
- FDA (2018) Bioanalytical method validation for industry guidance May 2018.
- https://www.fda.gov/drugs/guidance-compliance-regulatory-information/guidances-drugs
© by the Authors & Gavin Publishers. This is an Open Access Journal Article Published Under Attribution-Share Alike CC BY-SA: Creative Commons Attribution-Share Alike 4.0 International License. Read More About Open Access Policy.