Quality Assurance Phantom Testing of an Echo-Planar Diffusion-Weighted Sequence on a 3T Scanner
Muna Al-Mulla1*, Allison McGee2, Patrick Kenny3, Louise Rainford4
1Department
of Health Sciences, University of Kuwait, Kuwait
2University College
Dublin School of Medicine and Medical Science, Health Sciences Centre,
University College Dublin, Belfield, Dublin 4, Ireland
3Department
of Radiology,
Mater Misericordiae University Hospital, Level 2, The Whitty Building, North
Circular Road, Dublin 7. Ireland
4University
College Dublin School of Medicine and Medical Science, Health Sciences Centre,
University College Dublin, Belfield,
Dublin 4, Ireland
*Corresponding author: Muna Al-Mulla, Department of Health Sciences, University of Kuwait. Email: muna@HSC.EDU.KW
Received
Date: 10 December, 2018; Accepted Date: 10 January,2019; Published Date: 18 January, 2019
Citation: Al-Mulla M, McGee A, Kenny P, Rainford L (2019) Quality Assurance Phantom Testing of an Echo-Planar Diffusion-Weighted Sequence on a 3T Scanner. Adv Res Foot Ankle: ARFA-110. DOI: 10.29011/ ARFA-110.1000010.
Purpose: This
work investigated the behaviour of a high-field MR system relative to a range
of pulse sequence parameter changes and image artefacts resulting from image
acquisition employing an Echo-Planar Diffusion-Weighted (EPI-DW) sequence
modified for musculoskeletal imaging of the Achilles Tendon (AT).
Methods
and Materials: MR scanning was undertaken on a 3T
Philips Achieva MR scanner with an 8-channel foot/ ankle coil. Image quality
evaluation was based on images acquired using two phantoms: 1) an L-shaped foot/
ankle phantom containing Nickel Chloride (NiCl2)
in water and 2) a small cylindrical plastic phantom containing copper sulphate
(CuSO4)
in water. Quality control (QC) measurements were based on the American College
of Radiology (ACR) specifications. Signal-to-noise ratio (SNR), image intensity
uniformity, percentage ghosting, and geometric distortion were measured for
phantom images acquired using four sequences (T1W
SE, T2W
TSE, STIR, EPI-DW).
Results: The
performance of the EPI-DW sequence was tested according to ACR specifications
and compared to three standard pulse sequences routinely used for scanning of
the AT. The EPI-DW sequence met the ACR criteria for SNR and image intensity
uniformity, but failed to meet geometric distortion criteria. In terms of
percentage ghosting, artefacts were evident in the EPI-DW and ADC map images.
However, when quantified, these images remained within ACR specifications for
image ghosting.
Conclusion: When
modifying a pulse sequence for a new application, it is essential to understand
in advance the technical performance characteristics of the MR system and their
potential impact on resultant MR image quality. This study demonstrated that
the results of image quality testing revealed important findings that facilitated
further optimisation of the EPI-DW sequence prior to its application for
Achilles tendon scanning in human subjects.
1.
Introduction
Over the past few
years Magnetic Resonance Imaging (MRI) has rapidly developed in both clinical
practice and research. High field strength MR scanners (i.e., 3 Tesla) have
enabled new pulse sequences and advanced imaging techniques to be applied in
clinical practice, making MR image acquisition faster, diverse and more
sensitive to pathology [1,2].
High field systems have facilitated the application of a wider range of pulse
sequences in routine clinical practice rather than limiting them principally to
the research arena. In this context, the application of Diffusion-Weighted
Imaging (DWI) has expanded from an MR imaging technique used exclusively for
neuroimaging to applications in the body and musculoskeletal system [3].
DWI is a technique that measures the thermal motion of molecules within
tissues. When coupled with an Apparent Diffusion Coefficient (ADC) map, DWI can
be used to generate quantitative values representing pathological changes
associated with various diseases [4].
The principal application of DWI is in brain imaging for the assessment of
infarct and stroke-related pathology; however, more recently it is being
increasingly and routinely used for tumour staging, size measurements and
morphological characteristics [4].
However, with advanced sequences such as DWI, which is based on Echo-Planar
Imaging (EPI), it has become increasingly challenging to maintain high image
quality to facilitate accurate depiction of anatomy and pathological changes.
MR images generated using an EPI-DWI sequence may suffer from an inherent low
Signal-To-Noise Ratio (SNR) and can be very sensitive to several types of
artefacts such as blurring or motion; some of which may be sufficiently severe
to hinder the extraction of accurate quantitative Apparent Diffusion
Coefficient (ADC) values [5]. MR systems are
susceptible to technical problems due to the complex, interrelated demands on
the system to maintain high-performance imaging [6].
Artefacts (e.g. eddy current, motion, ghosting and chemical shift) in MRI can
be the result of several factors, including: magnetic field imperfections that
are not normally visible, scanner hardware or software limitation or
malfunction, or as a consequence of the characteristics of the subject
scanned [7]. Although techniques
have been developed to reduce these artefacts, not every artefact can be
avoided or minimized, with some being very unpredictable and difficult to
overcome [1]. Several authors have
acknowledged the complexity of those artefacts resulting from the use of
high-field systems to acquire certain types of advanced sequences, one of such
is the EPI-DWI discussed in this paper [2,7];
and have advocated the application of Quality Control (QC) testing in the
clinical setting specifically for this sequence. Standard QC tests can help
maintain MR system performance, facilitate the recognition of the source of
defects within MR images, and enable analysis and tuning of MR system
performance. In QC testing, the focus is on optimising MR system
characteristics such as spatial resolution (ghosting), linearity (geometric
distortion), homogeneity (image intensity uniformity), and signal
(signal-to-noise ratio) [8,9].
Routine QC phantom scanning can either be performed weekly and/ or monthly at
each site, and tends to be based on visual inspection of the MR images.
However, [10] has
indicated that this approach is insufficient for the EPI-DWI sequence, which is
prone to inconsistencies in measured values for some of the above indices of MR
image quality. Given the resultant difficulties evaluating artefact severity
across a number of EPI-DW images, stand-alone testing was recommended together
with quantitative evaluation of image quality indices This study aimed to
demonstrate the impact of performing MR phantom scanning and quantifying the
results using a computer-based approach for QC measurements in the development
of a new sequence for application in a clinical study. The technical
characteristics and artefacts associated with the acquisition of an EPI-DWI
sequence modified for ankle joint imaging images on a 3T MR scanner were
explored. These findings enabled parameter optimisation prior to application of
this sequence for Achilles tendon scanning on human subjects.
2.
Materials and Methods
A multi-step QC
testing process was undertaken to compare the performance of the EPI-DWI
sequence against a series of standard pulse sequences routinely used for
musculoskeletal MR imaging. The methodology was approved by the local Research
Ethics Committee. Four indices of MR image quality were evaluated using the
methodology for phantom scanning and evaluation proposed by the ACR (American
College of Radiology) [11]:
geometric distortion (aspect ratio), Signal-To-Noise Ratio (SNR), image
intensity uniformity, and percentage signal ghosting. Values for each of these
indices, as recommended by the ACR, are outlined in (Table
2).
2.1. MR
System and Phantom Specifications
Phantom MR imaging was
undertaken on a 3 Tesla Philips Achieve a scanner with the following gradient
characteristics: 40mT/m, 200mT/m/m sec
and incorporating an integrated 8-element, SENSE phased array
Radiofrequency (RF) coil dedicated for foot/ ankle imaging (Figure
1c). The parameters for the standard sequences: T1W
Spin Echo (SE), Turbo Spin Echo (TSE), Short Tau Inversion Recovery (STIR) and
the EPI-DW sequence tested are presented in Table
1. All phantom images were acquired in the axial
orientation.
The two phantoms
provided by the MR scanner vendor, were kept at room temperature prior to
scanning and had the following characteristics:
a) 31.8 mm/L Nickel
Chloride in water (NiCl2-H2O),
L-shaped to fit inside the dedicated foot/ ankle RF coil (Figure 1a);
b) 770 mg of Copper
Sulphate (CuSO4)
in water contained within a small, cylindrical plastic bottle (Fig
1b)
Axial images of the
NiCl2 phantom
were first acquired using, in order, a T1W SE, T2W FSE, STIR and EPI-DWI
sequence to test for geometric distortion (aspect ratio), Signal-To-Noise
Ratio (SNR), image intensity uniformity, and percentage signal ghosting. For consistency, each sequence was acquired
twice using same parameters (Table 1).
To correct for artefact resulting from using a diffusion-weighted sequence,
measurements were repeated using the CuSO4 phantom to
evaluate image intensity uniformity and SNR.
2.2. MR
system specifications and Image Acquisition
MR
Phantom Image Analysis
The acquired
DICOM-format MR images of the phantoms were processed and analysed using Matlab
(Math Works, Inc.) and according to the ACR specifications for aspect ratio
(geometric distortion), Signal-To-Noise Ratio (SNR), image intensity uniformity
and percentage signal ghosting (Table 2).
Geometric distortion of all sequences was determined by using aspect ratio
(e.g. measurement of width to height), then comparing the aspect ratio (true
dimension) of the curved-border of the L-shaped MR phantom to that of the
scanned images (observed measurement). The MR phantom and all sequence
measurements were taken at 90° to each other,
with mean height and length values recorded to provide the vertical and
horizontal extent of the target object within the (axial) MR images (Figure 2).
Once an aspect ratio was determined, the percent distortion was calculated
using the formula from (Table 2).
SNR was determined by
isolating the target object in the MR images using a threshold value derived
from the histogram (Figure 4) of the pixel value frequency. In
each case, a minimum pixel value between the distributions for object and
background was selected as the global threshold value [9].
A value for noise was estimated as the mean value of the background pixels,
excluding zero-value pixels. Using measurement tools available through Matlab
software on the standard PC used for quantitative phantom image analysis, the
interior 75% of the target was located using a morphological ‘close’ operation
and the mean pixel value of this region represented the mean signal intensity
of the phantom. The SNR was then obtained by dividing the standard deviation of
the outside ROI from the mean of the inside ROI (formula Table
2).
A measure of ghosting
ratio was derived according to ACR recommendations. The mean signal from the
phantom (object) was obtained from within a large target Region-Of-Interest
(ROI) that was more than 70% of the cross-sectional area of the phantom. Mean
signals were also taken from the image background in the frequency-encoding
direction (Top + Bottom) and in the phase-encoding direction (Left +
Right) (Figure 5). Areas of each rectangular region
outside the phantom target were approximately 10% of the size of the interior
of the target object. The ghosting ratio was calculated using these rectangular
regions located approximately halfway between the phantom object and image
boundaries in the respective directions using the formula in (Table
2).
Data from the Matlab
and OsiriX software were then transferred to the Statistical Package for Social
Sciences (IBM SPSS version 20.0 Inc.) to facilitate calculation of mean and
standard deviation values and histogram analysis of all four image quality indices
for the phantom MR images acquired using the T1-W SE, T2-W
TSE, STIR and EPI-DW sequences (Table 3).
ADC maps of quantitative diffusion measurements taken were generated using
OsiriX software for MAC, incorporating an added software plugin. Using its “Grow
Region”, 3D segmentation and “ROI Volume” functional options, mean and Standard
Deviation (SD) ADC values were measured from the EPI-DW scans (scan 1 and scan
2, as each sequence was acquired twice) (Tables
4,5).
3.
Results
To determine geometric
distortion, aspect ratio measurements for the NiCl2 L-shaped
phantom were calculated using the formula:
According to the ACR
criteria specifications, percentile values calculated from images acquired
using each of the pulse sequences is considered acceptable if they are < 5%.
All routine sequences T1W
SE (3.8%), T2W
TSE (4.5%), and STIR (5%) passed the specification criteria for geometric
distortion. While, the EPI-DW sequence; b=0 (14%), b=40 (14%), b=273 (12.5%),
b=800 (12.5%) failed to remain within the desired criteria, scoring higher than
the desired < 5% for acceptable geometric distortion.
In terms of image intensity uniformity, images acquired using the STIR and
DW-EPI sequences achieved the highest scores across the four sequences
evaluated. However, none of the sequences achieved the ACR recommended value of
90-100% image intensity uniformity when measured from the MR NiCl2 phantom
images. When this index of image quality was re-evaluated for the images
acquired using the CuSO4-based
phantom, the performance of the EPI-DWI sequence improved, achieving image
intensity uniformity values well within the ACR-recommended range: (Mean / SD)
b=0 (91.7 / 1.5); b=40 (90.3 / 1.3); b=273 (91.3 / 1.3), b=800 (89.2 / 1.12).
Relative to percentage signal ghosting, this was found not to be of
significance for the images acquired for all sequences evaluated. However,
images acquired using the EPI-DW sequence achieved the highest score for this
image quality index indicating a higher prevalence of ghosting artefact (Table
3). According to ACR the acceptance criteria for SNR
measurements could not be specified in general terms since the values are MR
imaging system specific and dependent on other factors such as: RF coil
characteristics, scanning conditions, phantom T1 and
T2 values
etc. However, variations were noted in the SNR values calculated for the
sequences tested, with images acquired using the T1W
SE sequence scoring highest (228.91),
and those acquired using the T2W TSE sequence
achieving the lowest score (16.6) (Table
3). The EPI-DW sequence performed well, achieving the
second highest score after the T1W SE sequence using
the NiCl2 phantom,
with b=0 images scoring higher than those for b=800 (Table
3). Conversely, the EPI-DW images scored the highest
measured SNR using the CuSO4 phantom
(b=0 [446.18], b=40 [440.35], b=273 [282.03],
b=800 [98.68]), compared to SNR measured using the NiCl2 phantom
(b=0 [75.35], b=40 [70.71], b=273 [61.23], b=800
[31.48]) (Table
4). According to Numano [12], the diffusion coefficient of pure water at 20°C is 2.023 × 10−3 mm2/s. In this work,
the ADC values calculated from the EPI-DW images are summarised in (Table 4).
ADC measurements for the phantom were found to be consistent with literature,
with minimal variation related to temperature and concentration of water to
Nickel Chloride of the MR phantom at the time of scanning [3,12].
4.
Discussion
The QC tests performed
provided a simple and comprehensive assessment of the performance of an EPI-DW
sequence modified for Achilles tendon scanning. All image quality related
characteristics of the acquired phantom MR images were quantitatively
evaluated.
4.1. Geometric
Distortion
ACR specifications for
acceptable geometric distortion suggest that the absolute value for percentage
geometric distortion calculated using the formula in (Table 2) should
not exceed 5% [11]. MR system performance for this study, based on measurements
of the MR phantom and MR phantom images was such that the standard T1W SE, T2W
TSE and STIR sequences were all within ACR specifications for geometric
distortion. However, when using the EPI-DW sequences, MR phantom images failed
to meet ACR specifications, scoring higher than the specified 5%. The scoring
of the images acquired using the EPI-DW sequence in terms of aspect ratio was
1.11 to 1.14, according to the different b-values used, while the actual aspect
ratio of the MR phantom was 1.3, indicating a scoring difference of 12-14% from
the true dimension, i.e., the true value. In general, the American Association
of Physicists in Medicine (AAPM) indicate that a failure in a geometric
distortion test can be attributed to gradient non-uniformity, as the linearity
of the gradient magnetic field is the principal hardware factor affecting
geometric accuracy [9]. However, as only the EPI-DW sequence failed the
ACR geometric distortion specification a more sequence-specific factor is
implicated. According to Ardekani & Sinha [13], Echo-Planar Imaging
(EPI) based diffusion-weighted image acquisition suffers from geometric
distortions due to both local magnetic field in homogeneities and eddy current
effects that arise from the large diffusion gradients required to obtain
diffusion-weighted images. Thus, the distortion artefact is related to the
composition of the DW sequence, specifically the EPI-based acceleration
technique. These authors [13] then further explain that when using
EPI techniques coupled with a DW sequence there are inevitable trade-off
factors inherent to EPI, of which the most noteworthy is the higher sensitivity
to susceptibility-induced magnetic field distortions. Literature recommends the
use of field map correction, short-axis readout, and parallel imaging
acceleration techniques to reduce geometric distortion artefacts for an
improved quality of the EPI-DW acquisition [14]. Identification of
geometric distortion in EPI-DW images is important, specifically for MSK MR
scanning because of the variation in tissue composition (e.g. bone, fat,
fluid), which makes identification of artefacts such as. motion, eddy current
and shimming failure complex [7].
4.2. Percent
Signal Ghosting
In this study, the
degree of image ghosting was quantified under the assumption that no motion
occurred within the MR phantom images as movement was reduced by using padding
during image acquisition, and MR system vibration was minimal. The ACR
specifies that for ghosting artefact to be image acceptable it should measure
less than or equal to 0.025. From (Table 3) it is evident that images
acquired using all the routine pulse sequences and the EPI-DW sequence acquired
at low b-values (b=0, 273, 400) passed the ACR criteria for signal ghosting.
However, images acquired using the EPI-DW sequence with a high b-value (b=800)
displayed minimal signal ghosting. According to Bammer et al [15],
different b-values can influence the degree of ghosting artifact, which was
evident from the results for the EPI-DW sequence. The use of a high b-value is
necessary to detect variations in pathology from minor tissue degradation to
complete destruction of the Achilles tendon [16]. Thus, it is essential to
understand the underlying reason for the ghosting artefact in order to implement
the correct technique to eliminate it. Since ghosting was more prominent in the
images acquired using the EPI-DW sequence at high b-values of 800, it was
concluded that this artifact occurred as a result of the influence of DW and
EPI in combination within the sequence. A number of steps can be implemented to
eliminate ghosting artefact related to the EPI-DW sequence several including:
the application of inner volume shims and more complete spectral fat
suppression (Spectral Attenuated Inversion Recovery SPIR or SPAIR), which are
possible to apply on the 3T scanners.
4.3. Image
Intensity Uniformity
According to
literature [17,9,11] the common causes of poor sequence performance
in an image intensity uniformity test include incorrect phantom positioning, ghosting,
and RF coil failure. According to ACR guidelines, the Percentage Intensity
Uniformity (PIU) for MR scanners operating at field strengths greater than 2T
is expected to be between 90-100%. However, if using water-filled phantoms at
field strengths at or above 3T, the dielectric and penetration effects are more
prevalent and a figure of less than 90% intensity uniformity is acceptable. In
this study, all sequences tested failed to meet the ACR specification for image
intensity uniformity. Sobol [18] has suggested a reason for this
based on a study in which five different 3T MR scanners were tested for image
intensity uniformity using the ACR acquisition and measurement protocol, but
failed to meet the required value. This may be related to the RF properties of
the ACR phantom due to limitations imposed by the electromagnetic physics and
the resonant RF frequency of 3T MRI systems. Furthermore, since the focus of
this research involved testing the quality of images acquired using an EPI-DW
sequence, a plastic cylindrical bottle filled with Copper Sulphate
solution was used to re-test the image intensity uniformity (Table 5).
EPI-DW images scores determined using the images from this CuSO4-based
phantom were within 90% of the ACR specification for percentage image intensity
uniformity, indicating poor compliance for images acquired using the phantom
comprising Nickel Chloride in solution. The AAPM indicate that water-filled
phantoms as described above are not optimal for acceptance QA testing of
high-field scanners due to RF penetration and dielectric effects that become
more pronounced with increasing frequency [9]. To avoid obtaining low
image intensity uniformity values for such high-field systems, Alecci et
al [19] recommend the use of oil-based QA phantoms, which minimise
the dielectric and penetration effects that occur at field strengths of 3T and
above.
4.4. Signal-to-Noise-Ratio
An important advantage
when scanning at high magnetic field strength is the increased MR signal
generated [20].
This increased signal is particularly useful when optimising SNR for DWI
sequences acquired at different b-values, and specifically high b-values, which
tend to reduce the inherent SNR [4].
However, this magnitude of SNR increase is not attained in practice because SNR
is influenced by several factors including: physiological noise, scanner
hardware characteristics, Radio-Frequency (RF) field inhomogeneity and
increased susceptibility effects [21].
Furthermore, SNR also depends on the metabolite concentration comprising the QA
phantom and the T1 /
T2 relaxation
times of the metabolites [22].
While changes in T1 relaxation
time with increasing field strength are well understood, the effects of the
higher magnetic field strength on T2 relaxation
times are not as predictable. Cochlin & Blamire [22] further
explain that T1 and
T2 are
largely independent of each other: T1 principally
determined by the amount of paramagnetic ions present and T2 primarily
a function of the MR phantom concentration. This is seen in a study performed
by Li & Mirowitz [23] in
which different phantom concentrates were used and in all scans T2 scored
less than T1.
Similarly, a study by Chien-Chuan et al [24] involving
QA testing of five different 3T scanners, found that SNR for T2W sequences in
all the tests performed was lower than for the T1W
sequences. For the STIR sequence Li & Mirowitz [23],
explain that STIR is robust at all field of strengths and may be less
vulnerable to field inhomogeneity, though it often produces lower SNR.
Furthermore, given the many MR system characteristics upon which SNR depends,
acceptance criteria for SNR cannot be specified in general terms since the
values will always be system specific [24].
This is evident in the results for this study, as consistent with literature
evidence, variation in SNR measurements for all four sequences was noted (Figure 6).
For further confirmation of SNR behaviour and its dependency on different
parameters was evident when the EPI-DW sequence was re-tested with a phantom
containing a Copper sulphate solution, demonstrating an increase in the
measured SNR (Table 5).
Li & Mirowitz [25] explained
this in terms of the inherent sensitivity of EPI to magnetic susceptibility
effects. Susceptibility artefacts may be mitigated through the use of parallel
imaging, which has been shown to be particularly successful when used in
association with EPI-DW sequences [26].
4.5. Apparent
Diffusion Coefficient
ADC represents a
valuable biomarker of disease; thus, ADC quantitative measurements are
necessary for validation and calibration of EPI-DW sequences [27].
According to Graessner [3], ADC is very dependent on temperature, with
pure water having a diffusion coefficient of 3 x 10-3mm2/s
at body temperature, which serves as a standard for MR scanners. However, the
phantoms used in this study did not contain pure water and the mean ambient
room temperature was 19°C. Changes in the calculated ADC
values may be attributed to these variations together with some ghosting
artifact that influences overall image uniformity.
5.
Conclusion
3T imaging
availability is increasing within healthcare systems. The advantages of such
high-field scanners have been well documented for several clinical applications
including neuro imaging, MR angiography and MR imaging of the small anatomical
structures comprising the joints. However, a number of image artefacts are more
prominent at 3T. Understanding their physical origin can help radiographers and
MR scientists to manage these artefacts through pulse sequence and image
protocol optimisation and testing in advance of commencing clinical MR imaging
of volunteer and/or patient subjects. Applying phantom testing of the EPI-DW
sequence to be used for Achilles tendon imaging as part of a subsequent
clinical study inferred important information regarding the behaviour of the 3T
MR scanner and the RF coil relative to ACR-defined measurements of MR image
quality.
6.
Advances in Knowledge
The results of this
study are of relevance for research and clinical testing of EPI-DW pulse
sequences with ADC map measurements on high-field 3T MR scanners. This study
facilitated the continuation to a study designed to measure qualitative DWI and
quantitative ADC measurement of the Achilles tendon and related pathology.
7.
Acknowledgement
Special thanks to, Mr. David Costello MSc. Medical Physicist
(Mater Misericordiae University Hospital), and, Ms. Annette White MR Clinical
Specialist Radiographer (Cappagh National Orthopaedic Hospital).
Figure 1: MR
phantoms; (a) L-shaped Nickel Chloride (NiCl2)
in water (b) Cylindrical bottle containing a solution of Copper Sulphate (CuSO4) in water, and the foot/ ankle RF coil (c) used for QC
testing.
Figure 2: The phantom is an L-shaped
square with calculations at 90°
angles (white arrows) undertaken using Matlab software to measure mean values
corresponding to the four boundaries to provide the vertical and horizontal
extent of the target object.
Percentage Image Uniformity (PIU) was calculated using the Region-Of-Interest
(ROI) measurement tool to determine the mean signal intensity of 70% within the
interior region of the NiCl2 phantom
target object (Table 2). Regions of maximum (SMax) and minimum (SMin)
signal intensity, typically 0.5% (1 cm2), were located within images acquired using each of
the sequences evaluated (Figure 3). Mean signal intensity
measurements (PIU) were calculated according this formula (Table 2).
Figure 3: An image demonstrating regions of maximum and minimum signal intensity.
The ROI has been placed at what was visually estimated to be the largest 1 cm2 dark area (blue arrow), the ROI was placed at
what was visually estimated to be the brightest 1 cm2 (red arrow), following
the ACR specification for calculation.
Figure 4: Images show the method used
to measure Signal-To-Noise Ratio (SNR) using Matlab software automated for SNR
calculation. The target object in the image was isolated using a threshold
value derived from the histogram (a) of pixel value frequency. The interior 75%
(b) of the target was located using a morphological ‘close’ operation and the
mean pixel value of this region represented signal value.
Figure 5: Image displaying a measure of
ghosting ratio, derived from the ACR approach, where the target object is 70%
of the interior of the phantom, and the rectangles outside this target located
top, bottom, left & right in the respective frequency- and phase-encoding
directions.
Sequence Parameter | TR (ms) | TE (ms) | Slice Thickness (mm) | NEX (Avg) | Number of Slices | Parallel Imaging | Matrix | |
T1-W SE | 400 | 12 | 3 | 4 | 15 | No | 256x320 | |
T2-W TSE | 4500 | 81 | 3 | 4 | 15 | No | 218x320 | |
STIR | 3800 | 46 | 3 | 4 | 15 | No | 269x384 | |
EPI-DW | 2030 | 30 | 3 | 12 | 15 | SENSE | 112x526 | |
Table 2: Criteria used for evaluating the phantom images obtained using the four pulse sequences mentioned in (Table 1).
Sequence | Geometric Distortion (Aspect Ratio) Mean ± (SD) | Percent-Signal Ghosting | Image Intensity Uniformity Mean ± (SD) | Signal-To-Noise Ratio Mean ± (SD) |
T2W Scan 1 | 1.25 ± (0.07) | 0.001± (0.004) | 51.43 ± (0.62) | 16.6 ± (0.061) |
T2W Scan 2 | 1.25 ± (0.07) | 0.001 ± (0.003) | 51.57 ± (0.84) | 16.60 ± (0.03) |
T1W Scan 1 | 1.26 ± (0.07) | 0.000 ± (0.000) | 56.13 ± (1.05) | 228.91 ± (26.93) |
T1W Scan 2 | 1.25 ± (0.07) | 0.000 ± (0.000) | 55.80 ± (0.73) | 233.50 ± (27.26) |
STIR Scan 1 | 1.24 ± (0.08) | 0.002 ± (0.003) | 60.28 ± (1.92) | 36.38 (2.39) |
STIR Scan 2 | 1.24 ± (0.08) | 0.003 ± (0.003) | 60.21 ± (1.33) | 45.98 ± (3.53) |
EPI-DW b=0 Scan 1 | 1.11 ± (0.16) | 0.011 ± (0.003) | 65.58 ± (1.95) | 79.17 ± (8.37) |
EPI-DW b=0 Scan 2 | 1.13 ± (0.18) | 0.013 ± (0.003) | 64.19 ± (2.51) | 71.53 ± (6.23) |
EPI-DW b=40 Scan 1 | 1.11 ± (0.16) | 0.011 ± (0.003) | 66.96 ± (2.24) | 76.91 ± (8.08) |
EPI-DW b=40 Scan 2 | 1.13 ± (0.17) | 0.013 ± (0.003) | 66.65 ± (2.64) | 64.51 ± (4.98) |
EPI-DW b=273 Scan 1 | 1.12 ± (0.17) | 0.010 ± (0.003) | 66.04 ± (2.26) | 65.61 ± (4.86) |
EPI-DW b=273 Scan 2 | 1.14 ± (0.17) | 0.013 ± (0.003) | 65.11 ± (2.34) | 56.86 ± (3.49) |
EPI-DW b=800 Scan 1 | 1.13 ± (0.17) | 0.010 ± (0.006) | 63.17 ± (1.90) | 34.66 ± (2.96) |
EPI-DW b=800 Scan 2 | 1.14 ± (0.17) | 0.013 ± (0.006) | 62.61 ± (2.72) | 28.31 ± (2.42) |
Table 3: QC Test Results for the L-shaped MR Phantom.
Sequence | Apparent Diffusion Coefficient (ADC) Values | ||
Nicl2 in Water Phantom Scan 1 Mean ± SD × 10−3 [Mm2/S] | Nicl2 in Water Phantom Scan 2 Mean ± SD × 10−3 [Mm2/S] | Cuso4 in Water Phantom Scan 1 Mean ± SD × 10−3 [Mm2/S] | |
EPI-DW b=0 | 2.1 ± (0.066) | 2.1 ± (0.059) | 2.0 ± (0.038) |
EPI-DW b=40 | 2.0 ± (0.373) | 2.0 ± (0.363) | 2.0 ± (0.047) |
EPI-DW b=273 | 2.0 ± (0.109) | 2.1 ± (0.109) | 2.0 ± (0.039) |
EPI-DW b=800 | 2.1 ± (0.060) | 2.1 ± (0.069) | 2.0 ± (0.048) |
Table 4: Apparent diffusion coefficient measurements for the NiCl2 and CuSO4 in water MR phantoms.
Sequence | Image Intensity Uniformity Cuso4 In Water Phantom Mean ± (SD) | Signal To Noise Ratio Cuso4 In Water Phantom Mean ± (SD) | Image Intensity Uniformity Nicl2 In Water Phantom Mean ± (SD) | Signal To Noise Ratio Nicl2 In Water Phantom Mean ± (SD) |
EPI-DW B=0 | 91.7 ± (1.5) | 446.18 ± (44.6) | 64.88 ± (2.23) | 75.35 ± (7.3) |
EPI-DW B=40 | 90.3 ± (1.3) | 440.35 ± (34.34) | 66.80 ± (2.44) | 70.71 ± (6.53) |
EPI-DW B=273 | 91.3 ± (1.3) | 282.03 ± (22.28) | 65.57 ± (2.3) | 61.23 ± (4.17) |
EPI-DW B=800 | 89.2 ± (1.1) | 98.68 ± (7.27) | 62.89 ± (2.31) | 31.48 ± (2.69) |
Table 5: Image quality indices for the EPI-DW sequence derived from images of Copper Sulphate (CuSO4) and Nickel Chloride (NiCl2) in water phantoms.
- Tanenbaum LN (2006) Clinical 3T MR imaging: mastering the challenges. Magn Reson Imaging Clin N Am 4: 1-15.
- Schmitt F, Grosu D, Mohr C, Purdy D, Salem K, et al. (2004) 3 Tesla MRI: successful results with higher field strengths. Radiologe 44: 31-47.
- Graessner J (2011) Frequently Asked Questions: Diffusion-Weighted Imaging (DWI). MAGNETOM Flash 84-87.
- Luna A, Ribes R, Soto JA (2012) Diffusion MRI outside the brain. Berlin: Springer.
- Liu Z, Wang Y, Gerig G, Gouttard S, Tao R, et al. (2010) Quality Control of Diffusion Weighted Images. Proc Soc Photo Opt Instrum Eng 11: 7628.
- Ihalainen TM, Lonnroth NT, Peltonen JI, Uusi-Simola JK, Timonen MH, et al. (2011) MRI quality assurance using the ACR phantom in a multi-unit imaging center. Acta Oncologica 50: 966-972.
- Le Bihan D, Poupon C, Amadon A, Lethimonnier F (2006) Artifacts and pitfalls in diffusion MRI. J Magn Reson Imaging 24: 478-488.
- Bushberg J (2012) The essential physics of medical imaging. Philadelphia; Wolters Kluwer Health/Lippincott Williams & Wilkins.
- Jackson EF, Bronskill MJ, Drost DJ, Och J, Pooley RA, et al. (2010) American Association of Physicists in Medicine. Acceptance Testing and Quality Assurance Procedures for Magnetic Resonance Imaging Facilities. AAPM REPORT NO. 100, American Association of Physicists in Medicine, Houston: American Association of Physicists in Medicine One Physics Ellipse.
- Delakis I, Moore EM, Leach MO, DE Wilde JP (2004) Developing a quality control protocol for diffusion imaging on a clinical MRI system. Phys Med Biol 49: 1490-1422.
- American College of Radiology. Phantom
test guidance for the ACR MRI accreditation program. Reston: American College
of Radiology, 2005.
- Numano T, Hyodo K, Nitta N, Hata J, Iwasaki N, et al. (2004) Apparent Diffusion Coefficient Mapping Using a Multi-Shot Spiral MRI Sequence of the Rat Brain. Open Journal of Radiology 4: 13-24.
- Ardekani S, Sinha U (2005) Geometric distortion correction of high-resolution 3 T diffusion tensor brain images. Magn Reson Med 54: 1163-1171.
- Chang HC, Chuang TC, Lin YR, Wang FN, Huang TY, et al. (2013) Correction of geometric distortion in Propeller echo planar imaging using a modified reversed gradient approach. Quant Imaging Med Surg 3: 73-81.
- Bammer R, Augustin M, Prokesch RW, Stollberger R, Fazekas F (2002) Diffusion-weighted imaging of the spinal cord: interleaved echo-planar imaging is superior to fast spin-echo. J Magn Reson Imaging 15: 364-373.
- Khoo MM, Tyler PA, Saifuddin A, Padhani AR (2011) Diffusion-weighted imaging (DWI) in musculoskeletal MRI: a critical review. Skeletal Radiol 40: 665-681.
- Chen CC, Wan YL, Wai YY, Liu HL (2004) Quality assurance of clinical MRI scanners using ACR MRI phantom: preliminary results. J Digit Imaging 17: 279-84.
- Sobol W (2006) Some Issues with Image Intensity Uniformity Test Performed On 3T MRI Scanners. Med Phys 33: 2209-2210.
- Alecci M, Collins C, Smith M, Jezzard P (2001) Radio Frequency Magnetic Field Mapping of a 3 Tesla Birdcage Coil: Experimental and Theoretical Dependence on Sample Properties. Magnetic Resonance in Medicine 46: 379-385.
- Schmitz BL, Aschoff AJ, Hoffmann MH, Gron G (2005) Advantages and Pitfalls in 3T MR Brain Imaging: A Pictorial Review. AJNR Am J Neuroradiol 26: 2229-2237.
- Chang K, Kamel I (2010) Abdominal imaging at 3T: challenges and solutions. Applied Radiology 39: 22-31.
- Cochlin L, Blamire A, Styles P (2003) Dependence of T1 and T2 on high field strengths in doped agarose gels; facilitating selection of composition for specific T1/T2 at relevant field. Proc. Intl. Soc. Mag. Reson 11: 885.
- Li T, Mirowitz SA (2002) Comparative study of fast MR imaging: quantitative analysis on image quality and efficiency among various time frames and contrast behaviours. Magn Reson Imaging 20: 471-478.
- National Electrical Manufacturers Association. Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging. 2008. National Electrical Manufacturers Association, Rosslyn (VA), USA: NEMA Standards Publications MS 1- 2008.
- Li T, Mirowitz SA (2003) Fast T2-weighted MR imaging: impact of variation in pulse sequence parameters on image quality and artifacts. Magn Reson Imaging 21: 745-753.
- Skare S, Newbould RD, Clayton DB, Albers GW, Nagle S, et al. (2007) Clinical multishot DW-EPI through parallel imaging with considerations of susceptibility, motion, and noise. Magn Reson Med 57: 881-890.
- Matsuya R, Kuroda M, Matsumoto Y, Kato H, Matsuzaki H, et al. (2009) A new phantom using polyethylene glycol as an apparent diffusion coefficient standard for MR imaging. Int J Oncol 35: 893-900.
© by the Authors & Gavin Publishers. This is an Open Access Journal Article Published Under Attribution-Share Alike CC BY-SA: Creative Commons Attribution-Share Alike 4.0 International License. Read More About Open Access Policy.